Instructions for Use
Leica ASP200 S -
Advanced Smart Processor
Vacuum
Tissue Processor
Leica ASP200 S, V 1.4 RevE, English – 03/2013
Order No.: 14 0480 80101, RevE
Always keep this manual with the instrument.
Read carefully before working with the instrument.
Vue de la page 0
1 2 3 4 5 6 ... 87 88

Résumé du contenu

Page 1 - Leica ASP200 S

Instructions for UseLeica ASP200 S -Advanced Smart ProcessorVacuumTissue ProcessorLeica ASP200 S, V 1.4 RevE, English – 03/2013Order No.: 14 0480 8010

Page 2

10Instructions for Use V 1.4, RevE – 03/20132. SafetyTransport and InstallationWarnings – Markings on the instrument itselfMarkings on the instrument

Page 3

11Leica ASP200 S2. SafetyWarnings – Operating the instrumentThe instrument may only be operated by trained laboratory personnel. It must only be op

Page 4 - Contents

12Instructions for Use V 1.4, RevE – 03/20132. SafetyWarnings – Handling reagentsWarnings – Cleaning and maintenanceTake care when handling solvents!A

Page 5

13Leica ASP200 S2.3 Safety features on the instrumentThe Leica ASP200 tissue processor is equipped with numerous safety func-tions and with sophistic

Page 6 - 1. Important Information

14Instructions for Use V 1.4, RevE – 03/20133.1 Overview – instrument components3. Instrument Components and Specifications2345136714151617181920212

Page 7 - 1. Important Information

15Leica ASP200 SOverview – instrument components1 - Basic instrument – processor module2 - Basic instrument – reagent module3 - Wax bath lids4

Page 8

16Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and Specifications3.2 Specific instrument options• Optionally usable Reagent M

Page 9

17Leica ASP200 S3. Instrument Components and Specifications3.3 Standard delivery - packing listTo prevent damage to the instrument or specimens, o

Page 10 - 2. Safety

18Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and SpecificationsPlease check all delivered parts against the packing list and

Page 11 - 2. Safety

19Leica ASP200 S3. Instrument Components and SpecificationsNominal supply voltages: Two factory-preset voltages (not user-adjustable):100 to 120 V

Page 13

20Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and SpecificationsRetortCapacity: max. 200 cassettesReagent volume: 3.5 l with v

Page 14

21Leica ASP200 S3.5 Compatible reagents3. Instrument Components and SpecificationsHardware and Software:• LargecolorLCDtouchscreen.• User-fri

Page 15 - Leica ASP200 S

22Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and SpecificationsThe following reagents may be used in the Leica ASP200 S:Fixat

Page 16

23Leica ASP200 S3.6.1 Cycle for changing reagentsReagent Week 1 Week 2 Week 3 Week 4Formalin At least 3 times a weekAt least 3 times a weekAt least 3

Page 17

24Instructions for Use V 1.4, RevE – 03/20134. Initial Operation4.1.1 Moving the instrumentAfter unpacking the instrument (see Unpack-ing instruction

Page 18

25Leica ASP200 S4. Instrument Setup4.2 Electrical connection• Checkthevoltagelabel(Fig. 3) on the rear of the instrument to ensure that the

Page 19 - 3.4 Technical Data

26Instructions for Use V 1.4, RevE – 03/2013• Connect the ASP to a grounded socket using the power cable (37).• Using the jumper cable (38), connect

Page 20

27Leica ASP200 S4. Instrument SetupFig. 5• Movetheinstrumenttoitsfinalsetuplocation.Active carbon filter• Unpacktheactivecarbonfilter

Page 21 - 3.5 Compatible reagents

28Instructions for Use V 1.4, RevE – 03/2013Fig. 8Inserting the volume replacement body2625272519Important!The instrument must be operated only when t

Page 22

29Leica ASP200 S4. Instrument SetupFig. 9Printer (optional)• Swingthedooroftheinstrumentconsole(6) upwards.• Connecttheprintertothe

Page 23

3Leica ASP200 SThe information, numerical data, notes and value judgments contained in this manual represent the current state of scientific knowledge

Page 24 - 4. Initial Operation

30Instructions for Use V 1.4, RevE – 03/20134. Initial OperationChanging the clip• Oncetheclipisinstalled,thehandle(4) of the speci-men baske

Page 25 - 4. Instrument Setup

31Leica ASP200 S4. Instrument SetupRemote alarmThis alarm is also external to the Leica ASP200 S.If installed, it might typically be connected to a

Page 26

32Instructions for Use V 1.4, RevE – 03/20134.7 Switching the instrument on4. Initial Operation• Afterbeingswitchedon,theinstrumentwilltake

Page 27

33Leica ASP200 SSYSTEM DIAGNOSTICS menuIf TOUCH HERE ... is touched during initialization (Fig. 14) and the pass-word for supervisor mode is en-tered,

Page 28

34Instructions for Use V 1.4, RevE – 03/20134.8 Touchscreen functionsThe Leica ASP200 S is programmed and operated via a color LCD touchscreen.The in

Page 29

35Leica ASP200 S4. Instrument Setup4.9 Checklist for initial operationParameter selectionSet warning values for the age of the carbon filter and t

Page 30

36Instructions for Use V 1.4, RevE – 03/20134.10 Switching off the instrumentIf the instrument must be completely switched off or disconnected from t

Page 31

37Leica ASP200 S5. Operation5.1 Setting up the instrument parametersThe SYSTEM SETUP screen is divided up into five selection fields:• Programo

Page 32

38Instructions for Use V 1.4, RevE – 03/2013PROGRAM OPTIONSSelect the options with which programs are executed here. Press a button to change the asso

Page 33

39Leica ASP200 SINSTRUMENTPress PARAFFIN BATH SET TEMP. to open the input screen (Fig. 30) for the wax bath temperature.Set the temperature according

Page 34

4Instructions for Use V 1.4, RevE – 03/2013Contents1. Important Information ...

Page 35

40Instructions for Use V 1.4, RevE – 03/2013SECURITYAccess rights to the instrument are managed here as user profiles.5. OperationSUPERVISOR MODE: En

Page 36

41Leica ASP200 S5. OperationIn the Leica ASP200 S, there is a lock mode to protect against operation by unauthorized per-sons.Enabled:The LOCK key

Page 37 - 5. Operation

42Instructions for Use V 1.4, RevE – 03/20135.1.3 INSTALLATION menu→→Use this menu to set the name of the instrument, the language of the user interf

Page 38 - 5. Operation

43Leica ASP200 SSetting the altitude of the installation location (in meters above sea level)This is important as it affects the actual pressure in th

Page 39

44Instructions for Use V 1.4, RevE – 03/2013Printing run logsPrinting is possible when the PRINT symbol is displayed.Configuring a printerOn the INSTA

Page 40

45Leica ASP200 S5.1.4 Editing the reagent listAdding new reagents1. Enter the reagent name and allocate the new reagent to the reagent group it belon

Page 41

46Instructions for Use V 1.4, RevE – 03/2013Entering / modifying reagent thresholdsIf warning thresholds are required for a certain reagent, enter the

Page 42

47Leica ASP200 SDeleting reagents• HighlightthereagenttobedeletedintheSET UP REAGENTS AND WARNING THRESHOLDS screen.• PressCLEAR.• Pres

Page 43

48Instructions for Use V 1.4, RevE – 03/20135.1.5 Viewing the program listYoucancreate:• Upto15infiltrationprograms• 3retortcleaningprogra

Page 44

49Leica ASP200 S5.1.6 Adding and/or modifying programsCreating a new program• Makesureyouareloggedonatsupervisorlevel.• IntheVIEW/EDIT P

Page 45

5Leica ASP200 SContents5.1.4 Editing the reagent list ...

Page 46

50Instructions for Use V 1.4, RevE – 03/20135. OperationEditing program steps• Toeditastep,highlightthecorrespondinglineandpresstherespec

Page 47

51Leica ASP200 SEditing program steps (continued)Setting a delay stepA delay step is a step that is extended to ensure that a program will finish at a

Page 48

52Instructions for Use V 1.4, RevE – 03/20135.1.7 FavoritesUp to five favorites can be configured in the ASP200 S. Favorites can be programmed so that

Page 49

53Leica ASP200 SAssigning a day for the end timeIf an end time has been assigned to a Favorite program, a specific weekday can also be set on which th

Page 50

54Instructions for Use V 1.4, RevE – 03/20135.1.8 Configuring the stationsAdding/changing a reagentSelect the desired station. Either by pressing the

Page 51

55Leica ASP200 S5.1.9 Reagent groupsFig. 53Changing the color of a reagent group• Highlightthelinecontainingthereagenttobemodified.• Presst

Page 52

56Instructions for Use V 1.4, RevE – 03/20135. Operation5.2 Reagent handling5.2.1 Draining/filling reagents (other than paraffin)1. Manually changin

Page 53

57Leica ASP200 S5. OperationRemote draining• Onthescreen,selectthereagentcontainer(No. 3) and the receptacle (External). Then press STATI

Page 54

58Instructions for Use V 1.4, RevE – 03/20135. Operation5.2.2 Replacing the paraffinDraining the paraffin• Connect the paraffindrainhose (36) to

Page 55

59Leica ASP200 SFilling with paraffin pellets5. OperationFig. 60• Filltheparaffinreservoirtothebrim(Fig. 60, 1) with paraffin pellets. This

Page 56

6Instructions for Use V 1.4, RevE – 03/20131. Important Information1.1 Symbols in the text and their meanings(5)ENTERWarnings appear in a gray box

Page 57

60Instructions for Use V 1.4, RevE – 03/2013Filling with liquid paraffinSoftware controlIf the instrument has been filled with liquid paraffin, the in

Page 58

61Leica ASP200 S5.3 Running programs There are two different windows from which a program can be started:In the Start screen, press the ALL

Page 59

62Instructions for Use V 1.4, RevE – 03/20135.3.1 Editing a program that has been called up. Fig. 65To modify a program before the start, press the ED

Page 60

63Leica ASP200 SThe RUN OPTIONS function displays the options with which the current program is started.• LEVEL TEST enabled: The lower level sensor

Page 61

64Instructions for Use V 1.4, RevE – 03/20135.3.2 Starting a programFig. 69↓→→5. OperationFig. 70• PressingSTART starts the program. Confirm the pro

Page 62

65Leica ASP200 S5. Operation5.3.3 Opening the retort during a program interruptionFig. 72 Fig. 73Fig. 71• After the signal tone stops, turn the le

Page 63

66Instructions for Use V 1.4, RevE – 03/20135. Operation5.3.4 Retort emergency release (continued)Fig. 74• Turn the retort lock further clockwise us

Page 64

67Leica ASP200 S5.4 Reagent statusThis window (Fig. 77) is used to display and update the current reagent status of the stations. In addition, reagen

Page 65

68Instructions for Use V 1.4, RevE – 03/20135. Operation5.5 System monitorThe SYSTEM MONITOR function provides important infor-mation about the statu

Page 66

69Leica ASP200 S5.6 The online helpThe Leica ASP200 S has an extensive help system that can be started from every main window.To obtain information a

Page 67

7Leica ASP200 S1.1 Symbols in the text and their meanings (continued)1. Important InformationSNREFSerial numberOrder No.Observe the Instructions f

Page 68

70Instructions for Use V 1.4, RevE – 03/20136. Troubleshooting6.1 General→For all error states, the instrument takes corrective measures to protect th

Page 69

71Leica ASP200 S6. Troubleshooting6.3 Troubleshooting→Use the SMART SCREEN functions to individually test operations such as fill, drain, pressuriz

Page 70 - 6. Troubleshooting

72Instructions for Use V 1.4, RevE – 03/20136. Troubleshooting6.4 Typical fill or drain problemsFill and drain problems may be due to several causes:

Page 71 - 6. Troubleshooting

73Leica ASP200 S7. Cleaning and Maintenance7.1 Clean programsThe cleaning programs are at the end of the list of all pro-grams (Fig. 45). To displ

Page 72

74Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceEditing steps• To edit a program step, highlight the corresponding line and p

Page 73 - 7.1 Clean programs

75Leica ASP200 S7.1.2 Paraffin cleaningIt is possible to clean used paraffin of solvent residues (xylene) in the Leica ASP200 S. To do so, select the

Page 74 - 7. Cleaning and Maintenance

76Instructions for Use V 1.4, RevE – 03/20137. Cleaning and maintenance7.1.3 Smart cleanThe smart cleaning prepares the ASP200 S for transport. Durin

Page 75 - 7.1.2 Paraffin cleaning

77Leica ASP200 S7.2 General cleaning stepsParaffin reservoirs7. Cleaning and MaintenanceRegularly inspect the collecting tray (16) for signs of le

Page 76 - 7. Cleaning and maintenance

78Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceReplacing the active carbon filter• Thelifeoftheactivecarbonfilterwill

Page 77 - 7.2 General cleaning steps

79Leica ASP200 SCleaning the retort lid• Removewaxfromtheinsideoftheretortlidwiththeplasticscraperprovided. Thoroughly remove all wax d

Page 78

8Instructions for Use V 1.4, RevE – 03/2013The instrument has been designed so that it is safe to use by the operator as well as for pro-cessing speci

Page 79

80Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceCleaning the surface of the infiltration moduleImportant in order to ensure th

Page 80

81Leica ASP200 SCleaning the reagent containers• Emptyandcleanthereagentcontainers. Using a bottle brush and a laboratory detergent in warm wa

Page 81

82Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceFig. 877.2.2 Periodic cleaning and maintenance (continued)44Fig. 88Fig. 89Use

Page 82

83Leica ASP200 S7. Cleaning and Maintenance7.3 Checklist for preventive maintenanceStep Daily Weekly Monthly every 3 monthsLubricate the O-rings o

Page 83

84Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceStep Daily Weekly Monthly All 3 monthsCheck the filter screens of the paraffin

Page 84

85Leica ASP200 S8. Warranty and ServiceWarrantyLeica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected

Page 85 - 8. Warranty and Service

86Instructions for Use V 1.4, RevE – 03/2013Appendix 1Notes on Specimen Preparation• Cut the tissue samples that are to be embedded to a size that fi

Page 87

www.LeicaBiosystems.comLeica Biosystems Nussloch GmbHHeidelberger Straße 17-19D- 69226 NusslochPhone: +49 6224 - 143 0Fax: +49 6224 - 143 268Web: www

Page 88

9Leica ASP200 S2.1 Safety notes2. SafetyThese Instructions for Use includes important information related to the operating safety and maintenance