Instructions for UseLeica ASP200 S -Advanced Smart ProcessorVacuumTissue ProcessorLeica ASP200 S, V 1.4 RevE, English – 03/2013Order No.: 14 0480 8010
10Instructions for Use V 1.4, RevE – 03/20132. SafetyTransport and InstallationWarnings – Markings on the instrument itselfMarkings on the instrument
11Leica ASP200 S2. SafetyWarnings – Operating the instrumentThe instrument may only be operated by trained laboratory personnel. It must only be op
12Instructions for Use V 1.4, RevE – 03/20132. SafetyWarnings – Handling reagentsWarnings – Cleaning and maintenanceTake care when handling solvents!A
13Leica ASP200 S2.3 Safety features on the instrumentThe Leica ASP200 tissue processor is equipped with numerous safety func-tions and with sophistic
14Instructions for Use V 1.4, RevE – 03/20133.1 Overview – instrument components3. Instrument Components and Specifications2345136714151617181920212
15Leica ASP200 SOverview – instrument components1 - Basic instrument – processor module2 - Basic instrument – reagent module3 - Wax bath lids4
16Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and Specifications3.2 Specific instrument options• Optionally usable Reagent M
17Leica ASP200 S3. Instrument Components and Specifications3.3 Standard delivery - packing listTo prevent damage to the instrument or specimens, o
18Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and SpecificationsPlease check all delivered parts against the packing list and
19Leica ASP200 S3. Instrument Components and SpecificationsNominal supply voltages: Two factory-preset voltages (not user-adjustable):100 to 120 V
20Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and SpecificationsRetortCapacity: max. 200 cassettesReagent volume: 3.5 l with v
21Leica ASP200 S3.5 Compatible reagents3. Instrument Components and SpecificationsHardware and Software:• LargecolorLCDtouchscreen.• User-fri
22Instructions for Use V 1.4, RevE – 03/20133. Instrument Components and SpecificationsThe following reagents may be used in the Leica ASP200 S:Fixat
23Leica ASP200 S3.6.1 Cycle for changing reagentsReagent Week 1 Week 2 Week 3 Week 4Formalin At least 3 times a weekAt least 3 times a weekAt least 3
24Instructions for Use V 1.4, RevE – 03/20134. Initial Operation4.1.1 Moving the instrumentAfter unpacking the instrument (see Unpack-ing instruction
25Leica ASP200 S4. Instrument Setup4.2 Electrical connection• Checkthevoltagelabel(Fig. 3) on the rear of the instrument to ensure that the
26Instructions for Use V 1.4, RevE – 03/2013• Connect the ASP to a grounded socket using the power cable (37).• Using the jumper cable (38), connect
27Leica ASP200 S4. Instrument SetupFig. 5• Movetheinstrumenttoitsfinalsetuplocation.Active carbon filter• Unpacktheactivecarbonfilter
28Instructions for Use V 1.4, RevE – 03/2013Fig. 8Inserting the volume replacement body2625272519Important!The instrument must be operated only when t
29Leica ASP200 S4. Instrument SetupFig. 9Printer (optional)• Swingthedooroftheinstrumentconsole(6) upwards.• Connecttheprintertothe
3Leica ASP200 SThe information, numerical data, notes and value judgments contained in this manual represent the current state of scientific knowledge
30Instructions for Use V 1.4, RevE – 03/20134. Initial OperationChanging the clip• Oncetheclipisinstalled,thehandle(4) of the speci-men baske
31Leica ASP200 S4. Instrument SetupRemote alarmThis alarm is also external to the Leica ASP200 S.If installed, it might typically be connected to a
32Instructions for Use V 1.4, RevE – 03/20134.7 Switching the instrument on4. Initial Operation• Afterbeingswitchedon,theinstrumentwilltake
33Leica ASP200 SSYSTEM DIAGNOSTICS menuIf TOUCH HERE ... is touched during initialization (Fig. 14) and the pass-word for supervisor mode is en-tered,
34Instructions for Use V 1.4, RevE – 03/20134.8 Touchscreen functionsThe Leica ASP200 S is programmed and operated via a color LCD touchscreen.The in
35Leica ASP200 S4. Instrument Setup4.9 Checklist for initial operationParameter selectionSet warning values for the age of the carbon filter and t
36Instructions for Use V 1.4, RevE – 03/20134.10 Switching off the instrumentIf the instrument must be completely switched off or disconnected from t
37Leica ASP200 S5. Operation5.1 Setting up the instrument parametersThe SYSTEM SETUP screen is divided up into five selection fields:• Programo
38Instructions for Use V 1.4, RevE – 03/2013PROGRAM OPTIONSSelect the options with which programs are executed here. Press a button to change the asso
39Leica ASP200 SINSTRUMENTPress PARAFFIN BATH SET TEMP. to open the input screen (Fig. 30) for the wax bath temperature.Set the temperature according
4Instructions for Use V 1.4, RevE – 03/2013Contents1. Important Information ...
40Instructions for Use V 1.4, RevE – 03/2013SECURITYAccess rights to the instrument are managed here as user profiles.5. OperationSUPERVISOR MODE: En
41Leica ASP200 S5. OperationIn the Leica ASP200 S, there is a lock mode to protect against operation by unauthorized per-sons.Enabled:The LOCK key
42Instructions for Use V 1.4, RevE – 03/20135.1.3 INSTALLATION menu→→Use this menu to set the name of the instrument, the language of the user interf
43Leica ASP200 SSetting the altitude of the installation location (in meters above sea level)This is important as it affects the actual pressure in th
44Instructions for Use V 1.4, RevE – 03/2013Printing run logsPrinting is possible when the PRINT symbol is displayed.Configuring a printerOn the INSTA
45Leica ASP200 S5.1.4 Editing the reagent listAdding new reagents1. Enter the reagent name and allocate the new reagent to the reagent group it belon
46Instructions for Use V 1.4, RevE – 03/2013Entering / modifying reagent thresholdsIf warning thresholds are required for a certain reagent, enter the
47Leica ASP200 SDeleting reagents• HighlightthereagenttobedeletedintheSET UP REAGENTS AND WARNING THRESHOLDS screen.• PressCLEAR.• Pres
48Instructions for Use V 1.4, RevE – 03/20135.1.5 Viewing the program listYoucancreate:• Upto15infiltrationprograms• 3retortcleaningprogra
49Leica ASP200 S5.1.6 Adding and/or modifying programsCreating a new program• Makesureyouareloggedonatsupervisorlevel.• IntheVIEW/EDIT P
5Leica ASP200 SContents5.1.4 Editing the reagent list ...
50Instructions for Use V 1.4, RevE – 03/20135. OperationEditing program steps• Toeditastep,highlightthecorrespondinglineandpresstherespec
51Leica ASP200 SEditing program steps (continued)Setting a delay stepA delay step is a step that is extended to ensure that a program will finish at a
52Instructions for Use V 1.4, RevE – 03/20135.1.7 FavoritesUp to five favorites can be configured in the ASP200 S. Favorites can be programmed so that
53Leica ASP200 SAssigning a day for the end timeIf an end time has been assigned to a Favorite program, a specific weekday can also be set on which th
54Instructions for Use V 1.4, RevE – 03/20135.1.8 Configuring the stationsAdding/changing a reagentSelect the desired station. Either by pressing the
55Leica ASP200 S5.1.9 Reagent groupsFig. 53Changing the color of a reagent group• Highlightthelinecontainingthereagenttobemodified.• Presst
56Instructions for Use V 1.4, RevE – 03/20135. Operation5.2 Reagent handling5.2.1 Draining/filling reagents (other than paraffin)1. Manually changin
57Leica ASP200 S5. OperationRemote draining• Onthescreen,selectthereagentcontainer(No. 3) and the receptacle (External). Then press STATI
58Instructions for Use V 1.4, RevE – 03/20135. Operation5.2.2 Replacing the paraffinDraining the paraffin• Connect the paraffindrainhose (36) to
59Leica ASP200 SFilling with paraffin pellets5. OperationFig. 60• Filltheparaffinreservoirtothebrim(Fig. 60, 1) with paraffin pellets. This
6Instructions for Use V 1.4, RevE – 03/20131. Important Information1.1 Symbols in the text and their meanings(5)ENTERWarnings appear in a gray box
60Instructions for Use V 1.4, RevE – 03/2013Filling with liquid paraffinSoftware controlIf the instrument has been filled with liquid paraffin, the in
61Leica ASP200 S5.3 Running programs There are two different windows from which a program can be started:In the Start screen, press the ALL
62Instructions for Use V 1.4, RevE – 03/20135.3.1 Editing a program that has been called up. Fig. 65To modify a program before the start, press the ED
63Leica ASP200 SThe RUN OPTIONS function displays the options with which the current program is started.• LEVEL TEST enabled: The lower level sensor
64Instructions for Use V 1.4, RevE – 03/20135.3.2 Starting a programFig. 69↓→→5. OperationFig. 70• PressingSTART starts the program. Confirm the pro
65Leica ASP200 S5. Operation5.3.3 Opening the retort during a program interruptionFig. 72 Fig. 73Fig. 71• After the signal tone stops, turn the le
66Instructions for Use V 1.4, RevE – 03/20135. Operation5.3.4 Retort emergency release (continued)Fig. 74• Turn the retort lock further clockwise us
67Leica ASP200 S5.4 Reagent statusThis window (Fig. 77) is used to display and update the current reagent status of the stations. In addition, reagen
68Instructions for Use V 1.4, RevE – 03/20135. Operation5.5 System monitorThe SYSTEM MONITOR function provides important infor-mation about the statu
69Leica ASP200 S5.6 The online helpThe Leica ASP200 S has an extensive help system that can be started from every main window.To obtain information a
7Leica ASP200 S1.1 Symbols in the text and their meanings (continued)1. Important InformationSNREFSerial numberOrder No.Observe the Instructions f
70Instructions for Use V 1.4, RevE – 03/20136. Troubleshooting6.1 General→For all error states, the instrument takes corrective measures to protect th
71Leica ASP200 S6. Troubleshooting6.3 Troubleshooting→Use the SMART SCREEN functions to individually test operations such as fill, drain, pressuriz
72Instructions for Use V 1.4, RevE – 03/20136. Troubleshooting6.4 Typical fill or drain problemsFill and drain problems may be due to several causes:
73Leica ASP200 S7. Cleaning and Maintenance7.1 Clean programsThe cleaning programs are at the end of the list of all pro-grams (Fig. 45). To displ
74Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceEditing steps• To edit a program step, highlight the corresponding line and p
75Leica ASP200 S7.1.2 Paraffin cleaningIt is possible to clean used paraffin of solvent residues (xylene) in the Leica ASP200 S. To do so, select the
76Instructions for Use V 1.4, RevE – 03/20137. Cleaning and maintenance7.1.3 Smart cleanThe smart cleaning prepares the ASP200 S for transport. Durin
77Leica ASP200 S7.2 General cleaning stepsParaffin reservoirs7. Cleaning and MaintenanceRegularly inspect the collecting tray (16) for signs of le
78Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceReplacing the active carbon filter• Thelifeoftheactivecarbonfilterwill
79Leica ASP200 SCleaning the retort lid• Removewaxfromtheinsideoftheretortlidwiththeplasticscraperprovided. Thoroughly remove all wax d
8Instructions for Use V 1.4, RevE – 03/2013The instrument has been designed so that it is safe to use by the operator as well as for pro-cessing speci
80Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceCleaning the surface of the infiltration moduleImportant in order to ensure th
81Leica ASP200 SCleaning the reagent containers• Emptyandcleanthereagentcontainers. Using a bottle brush and a laboratory detergent in warm wa
82Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceFig. 877.2.2 Periodic cleaning and maintenance (continued)44Fig. 88Fig. 89Use
83Leica ASP200 S7. Cleaning and Maintenance7.3 Checklist for preventive maintenanceStep Daily Weekly Monthly every 3 monthsLubricate the O-rings o
84Instructions for Use V 1.4, RevE – 03/20137. Cleaning and MaintenanceStep Daily Weekly Monthly All 3 monthsCheck the filter screens of the paraffin
85Leica ASP200 S8. Warranty and ServiceWarrantyLeica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected
86Instructions for Use V 1.4, RevE – 03/2013Appendix 1Notes on Specimen Preparation• Cut the tissue samples that are to be embedded to a size that fi
www.LeicaBiosystems.comLeica Biosystems Nussloch GmbHHeidelberger Straße 17-19D- 69226 NusslochPhone: +49 6224 - 143 0Fax: +49 6224 - 143 268Web: www
9Leica ASP200 S2.1 Safety notes2. SafetyThese Instructions for Use includes important information related to the operating safety and maintenance