Leica-biosystems SM2010 R Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Leica-biosystems SM2010 R. Leica Biosystems SM2010 R User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 46
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Instructions for Use
Leica SM2010 R
Sliding Microtome
Instructions for Use
Leica SM2010 R, V 1.4 RevC – 09/2013, English
Order No.: 14 0508 80101 RevC
Always keep this manual with the instrument.
Read carefully before working with the instrument.
Vue de la page 0
1 2 3 4 5 6 ... 45 46

Résumé du contenu

Page 1 - Leica SM2010 R

Instructions for UseLeica SM2010 RSliding MicrotomeInstructions for UseLeica SM2010 R, V 1.4 RevC – 09/2013, EnglishOrder No.: 14 0508 80101 RevCAlway

Page 2

8Instructions for Use V 1.4 RevC – 09/20133. Instrument components and specifications3.2 Technical dataGeneral Approvals: The instrument-speci

Page 3

9Leica SM2010 R3. Instrument Components and Specifications3.3 Instrument specifications• TheLeicaSM2010Risamanuallyoperatedslidingmicrot

Page 4 - Table of contents

10Instructions for Use V 1.4 RevC – 09/20134. Instrument Setup4.2 Site requirements• Stable,vibration-freelaboratorytablewithhorizontal,flat

Page 5 - 1. Important Information

11Leica SM2010 R4.3 Unpacking• Tolifttheinstrumentfromthebox,holditonthe left and right of the housing (Fig. 5), lift it out of the foam cu

Page 6 - 2. Safety

12Instructions for Use V 1.4 RevC – 09/20134. Instrument Setup2526Horizontal alignmentFor safe and accurate work, it is important that all instrument

Page 7 - 2. Safety

13Leica SM2010 R4.5 Assembling the knife holder SN• Locktheknifesledge(11) in place using the locking knob (23) so that it cannot be moved.• Th

Page 8

14Instructions for Use V 1.4 RevC – 09/2013Fastening the knife holder• Presstheclampinglever (66) of the knife holder (67) downwards and screw th

Page 9

15Leica SM2010 RFig. 13Setting the clearance angle (Fig. 13)• Unscrewtheknurledheadscrew(69).• Settherequiredanglewiththeadjustinglever

Page 10 - 3.2 Technical data

16Instructions for Use V 1.4 RevC – 09/20134.6 Assembling the blade holder SE4. Instrument SetupMounting the grooved plate• Locktheknifesledge(

Page 11

17Leica SM2010 R4. Instrument SetupInserting the blade holder SE• Pushthebladeholderfullytothebacksuchthat the T-piece can be inserted in

Page 13 - 4. Instrument Setup

18Instructions for Use V 1.4 RevC – 09/20134. Instrument SetupInclination (declination) of the blade holder SE• UnscrewtheAllenscrew(52) at the

Page 14

19Leica SM2010 RAdjusting the clearance angle and declination• Theindexmarks(0°to10°)foradjustingtheclearance angle (43.1) are on the right

Page 15

20Instructions for Use V 1.4 RevC – 09/20134. Instrument SetupThe object orientation allows for simple position correction of the specimen surface wh

Page 16

21Leica SM2010 R5. Operation5.1 Operating elements and their functionsThe section thickness is set by turning the adjust-ing knob (33) on the left

Page 17

22Instructions for Use V 1.4 RevC – 09/20135. Operation5.1.3 Manual feedThe lever (35) for the manual feed is at the front right of the instrument.•

Page 18

23Leica SM2010 R5. Operation5.1.5 Directional fixture for specimen clampsOrienting the specimen• Toloosen the clamp, rotatethe eccentricl

Page 19

24Instructions for Use V 1.4 RevC – 09/20135. Operation5.2 Clamping the specimen in the universal cassette clamp (UCC)23• Movethecassetteclamp(4

Page 20

25Leica SM2010 R5. Operation5.3 Clamping a disposable bladeInsert the blade (Fig. 28)• Push the knife guard (14) towards the right and push t

Page 21

26Instructions for Use V 1.4 RevC – 09/20135.4 Replacing the pressure plateThe blade holder SE can be converted from use of narrow blades to use of w

Page 22

27Leica SM2010 R• Locktheknifesledge(11) in place using the locking knob (23).• Pleasemakesurethattheknifeholderis firmly clamped using t

Page 23 - 5. Operation

1Leica SM2010 RLeica Biosystems Nussloch GmbHHeidelberger Str. 17 - 19D –69226 NusslochGermanyPhone: +49 62 24 143–0Fax: +49 6224 143–268 Interne

Page 24 - 5. Operation

28Instructions for Use V 1.4 RevC – 09/2013Cutting into the specimen (trimming)For trimming, the specimen feed can be dis-engaged either by turning th

Page 25

29Leica SM2010 R5.7 Changing the specimen or interrupting sectioning5.8 Finishing daily routine• Raisethespecimentothelowerendpositionbytu

Page 26

30Instructions for Use V 1.4 RevC – 09/20136.1 Cleaning the instrument6. Cleaning and maintenanceBefore each cleaning carry out the following prepar

Page 27

31Leica SM2010 RThe instrument is basically maintenance-free. To ensure trouble-free operation of the instrument over a long period of time, the follo

Page 28

32Instructions for Use V 1.4 RevC – 09/20137.1 Ordering informationSupermega cassette clamp with adapter, silver ...

Page 29

33Leica SM2010 RSupermega cassette clamp,with adapter, silverfor mounting in the quick change system for specimen imagingMaximum specimen size: 75 x 5

Page 30

34Instructions for Use V 1.4 RevC – 09/2013Standard specimen clamp,with adapter, silverfor mounting in the quick change system for specimen imagingMax

Page 31

35Leica SM2010 R7. Optional AccessoriesBlade holder SE, assemblyEasily converted from a low-profile to high-profile blade holder by switching the p

Page 32 - 6. Cleaning and maintenance

36Instructions for Use V 1.4 RevC – 09/20137. Optional accessoriesKnife holder SN, assemblyfor resharpenable knives or blade rails. Blade holder decl

Page 33 - 6.2 Maintenance instructions

37Leica SM2010 RPlastic knife guard,for blade railsOrder No. ... 14 0368 337727. Optional AccessoriesFig.

Page 34 - 7. Optional accessories

2Instructions for Use V 1.4 RevC – 09/2013Table of contents1. Important Information ...

Page 35 - 7. Optional Accessories

38Instructions for Use V 1.4 RevC – 09/2013Leica low-profile disposable blades - Type 819Length 80 mm, height 8 mm1 package of 50 piecesOrder No. ...

Page 36

39Leica SM2010 R8. TroubleshootingIn the following table there is a list of the most common problems which can arise while work-ing with the instru

Page 37

40Instructions for Use V 1.4 RevC – 09/2013WarrantyLeica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected

Page 38

41Leica SM2010 R10. Notes

Page 39

42Instructions for Use V 1.4 RevC – 09/201311. Decontamination Certificate (Master)Leica Biosystems Nussloch GmbHHeidelberger Str. 17-1969226 Nussloc

Page 40

43Leica SM2010 RLeica Biosystems Nussloch GmbHHeidelberger Str. 17-1969226 Nussloch, GermanyPhone: ++49 (0) 6224 143 0Fax: ++49 (0) 6224 143 268www.

Page 41 - 8. Troubleshooting

www.LeicaBiosystems.comLeica Biosystems Nussloch GmbHHeidelberger Straße 17–19D– 69226 NusslochPhone: +49 - (0) 6224 - 143 0Fax: +49 - (0) 6224 - 143

Page 42 - 9. Warranty and Service

3Leica SM2010 RSymbols in the text and their meanings1. Important Information(5)Intended use of instrumentThe Leica SM2010 R is a manually operated

Page 43 - 10. Notes

4Instructions for Use V 1.4 RevC – 09/20132. Safety2.1 Safety notesThese Instructions for Use include important information related to the operating

Page 44 - Decontamination Certificate

5Leica SM2010 R2. SafetyWarnings - Safety instructions / warning labels attached to the instrumentWarnings - Transport and installationWarnings - W

Page 45

6Instructions for Use V 1.4 RevC – 09/20132. Safety2.3 Integrated safety devicesThe instrument is equipped with the following safety devices:• Knife

Page 46

7Leica SM2010 R3. Instrument Components and SpecificationsSpecimen cylinderRemovable section waste tray3.1 Overview - instrument componentsKnife

Commentaires sur ces manuels

Pas de commentaire